Endothelin-1 (ET-1) is a 21-amino-acid vasoconstrictor peptide produced by vascular endothelial cells and widely implicated in coronary artery disease (CAD). Elevated plasma ET-1 concentrations have been associated with disease severity and with coronary microvascular dysfunction in people with CAD, and the peptide has been a pharmacodynamic target of interest since the first endothelin receptor antagonists entered clinical trials in the 1990s. Yet measuring ET-1 in plasma presents a technical challenge: the biologically active form has a characteristically short plasma half-life, making reliable quantification dependent on careful sample handling and assay specificity.
Enzyme-linked immunosorbent assay (ELISA) is the established clinical method, but it is inherently single-plex and consumes sample volume that is difficult to spare when coronary artery blood comes from invasive intracoronary catheterization. Whether a high-throughput proteomics platform can measure ET-1 as reliably as ELISA, in the same low-volume coronary samples, had lacked peer-reviewed data until now.
A new study, published in the Canadian Journal of Physiology and Pharmacology in collaboration with researchers at the Department of Cardiology, Royal Papworth Hospital and the Department of Medicine, University of Cambridge, addresses that gap. The paper reports a significant correlation between ET-1 measurements generated by Olink's proximity extension assay (PEA) technology and those from an established ELISA-based method, using intracoronary plasma samples contributed by PlaqueTec.
How PEA detects thousands of proteins from a single microliter
The Olink Explore 3072 platform detects proteins through PEA chemistry, where each target protein is simultaneously recognized by a pair of matched antibodies, each conjugated to a unique DNA oligonucleotide. When both antibodies bind their target, the oligonucleotides hybridize and generate a new DNA amplicon that barcodes the protein for downstream quantification by next-generation sequencing (NGS). Per Olink's product specifications, the Explore 3072 panel covers approximately 3,000 proteins from approximately 1µL of plasma per panel, providing specificity comparable to sandwich immunoassay at a far greater multiplexing depth.
Low sample volume makes PEA particularly relevant for intracoronary biomarker work. Coronary artery blood, collected through invasive catheterization, is available in limited quantities; any assay requiring tens to hundreds of microliters per analyte quickly becomes impractical at scale. ET-1's short plasma half-life, small molecular size, and low circulating concentrations made it a demanding test case for confirming PEA concordance with a reference immunoassay.
Olink PEA vs. ELISA: what the assay comparison showed
PlaqueTec's BIOPATTERN trial (NCT06259019) is a multicenter study of up to 300 people with established CAD who are undergoing coronary angiography at specialist National Health Service (NHS) cardiac centers across England. The trial's Liquid Biopsy System (LBS) is a coronary artery catheter that extracts blood simultaneously from multiple sites along a diseased coronary vessel, collecting intracoronary plasma fractions that reflect local protein gradients undetectable in peripheral venous blood. ET-1 measurements compared in this study came from those LBS samples, testing both the Olink Explore 3072 PEA platform and a validated clinical ELISA.
Taken together, the results confirm that PEA can do what ELISA cannot in this context: accurately detect active ET-1 from a difficult, low-volume sample type while remaining scalable to multiplex biomarker panels.
- Assay concordance: Significant correlation between Olink Explore 3072 PEA and ELISA for biologically active ET-1 in intracoronary plasma
- Isoform specificity: Active ET-1 detected selectively, rather than the larger, lower-activity precursor big endothelin-1
- Sample type: Intracoronary plasma from the PlaqueTec LBS, more anatomically proximate to disease than peripheral venous blood
- Trial relevance: Results support ET-1 inclusion in multiplex Olink panels for biomarker profiling where sample volumes are limited
UK Biobank proteomic data using Olink assays have previously shown ET-1 levels to be elevated in participants with pulmonary arterial hypertension (PAH), linked to worsening vascular resistance and disease severity. This history reinforces ET-1's value as a quantifiable signal across vascular pathologies and supports its inclusion in multiplex cardiovascular panels.
Diane Proudfoot, Chief Scientific Officer of PlaqueTec and co-author, said the validation directly supports interpretation of BIOPATTERN's intracoronary protein data. "Knowing that the Olink assay genuinely reflects biologically active ET-1 is really important for us," Proudfoot said. "It gives us confidence that the protein signals we're seeing in coronary blood samples from our BIOPATTERN trial are meaningful, and that we can trust the data as we work to understand the biology of atherosclerotic plaques and identify new ways to help patients with cardiovascular disease."
Not all endothelin-1 measurements are equal
ET-1 begins as a 212-amino-acid prepropeptide, processed first to big ET-1, a 38-amino-acid lower-activity intermediate, then cleaved further to the mature 21-amino-acid active peptide by endothelin-converting enzymes (ECEs). Big ET-1 persists longer in plasma but retains measurable biological activity and converts further to ET-1 in circulation, so it is not a fully inert surrogate. Assays that detect total endothelin immunoreactivity without isoform selectivity can conflate active ET-1 with big ET-1 and the inactive ET-2 and ET-3 isoforms, inflating apparent ET-1 concentrations and obscuring the biologically meaningful signal.
Dual-antibody binding blocks isoform cross-reactivity at the source
PEA's dual-antibody design addresses cross-reactivity directly. Both antibodies in the PEA pair must bind their target simultaneously before an amplicon generates; dual epitope occupancy creates a structural barrier that single-antibody ELISA formats cannot replicate in the same way. The concordance between Olink PEA and the reference ELISA for the active ET-1 isoform specifically provides analytical evidence that PEA can resolve this distinction in intracoronary plasma.
Where ELISA reaches its limits, multiplex proteomics steps in
ELISA is the gold standard for validated clinical analytes, with well-characterized performance, regulatory acceptance, and minimal infrastructure requirements. Its practical limitation is proportional scaling: every additional protein target requires more sample volume and more assay time. Of the four approaches compared below, only PEA combines high-multiplex protein coverage with microliter-scale sample input, which makes intracoronary proteomics feasible.
| Measurement approach | Target scope | Sample volume | Regulatory status | Typical application |
|---|---|---|---|---|
| Sandwich ELISA (clinical) | Single protein per assay | Typically tens to hundreds of µL per analyte | CE-marked or FDA-cleared for specific targets | Clinical diagnostics; single-biomarker pharmacodynamic studies |
| Olink PEA (Explore 3072) | ~3,000 proteins per run (manufacturer specifications) | ~1µL per 384-plex panel (manufacturer specifications) | Research use; not IVD-cleared for individual targets | Discovery proteomics; trial biomarker profiling; population studies |
| Label-free mass spectrometry | Thousands of proteins; untargeted | Variable; typically >50µL | Research use | Unbiased discovery; lower sensitivity for low-abundance peptides |
| Bead-based multiplex immunoassay | Tens of proteins per panel | Tens of µL per sample | Research use; CE-marked kits for select panels | Focused multi-protein panels in clinical studies |
Does validated ET-1 measurement open new doors for CAD drug development?
Beyond its role as a biomarker, ET-1 is a pharmacodynamically active molecule central to vascular drug development. Endothelin receptor type A (ETA) antagonists have established clinical profiles in pulmonary hypertension and are under active investigation in conditions such as microvascular angina, where elevated ET-1 has been linked to microvascular dysfunction in people with CAD.
Anthony Davenport, Professor of Cardiovascular Pharmacology at the University of Cambridge and co-author on the study, said the findings extend the platform's utility into the endothelin field. "Detrimental increases in plasma ET-1 are a key biomarker identifying diseases for successfully translating ET receptor antagonists into the clinic," Davenport said. "In clinical trials, increases in circulating ET-1 can also be exploited as a pharmacodynamic biomarker, providing evidence for receptor blockade and target engagement by ET antagonists, optimal dose selection and patient stratification."
Confirming that Olink PEA accurately captures active ET-1 from coronary samples means BIOPATTERN's dataset can generate pharmacodynamically interpretable signals relevant to patient endotyping and therapeutic target identification. A landmark endothelin pharmacology review in Pharmacological Reviews documents ET-1's role across pulmonary hypertension, hypertension, and CAD, framing this assay validation within a broader effort to translate ET-1 biology into clinical interventions.
A validated proteomics tool for the hardest samples in cardiovascular research
Validating Olink PEA for active ET-1 in intracoronary plasma confirms that low-volume, high-throughput proteomics applies reliably to one of the most analytically challenging cardiovascular sample types. BIOPATTERN's intracoronary sampling design, capturing site-of-disease protein gradients unavailable in peripheral blood, now has a validated multiplex proteomics tool for a clinically and pharmacologically important target. As the trial continues to enroll participants and build the BioCarta intracoronary protein database, ET-1 profiling by Olink PEA stands to contribute to patient endotyping and novel drug target identification in coronary atherosclerosis.
This article is based on a press release issued by PlaqueTec and was produced under Drug Discovery News' AI Editorial Guidelines.